Measuring ranges:
0-200.0mΩ in steps of 100μΩ
0-2000mΩ in steps of 1mΩ
0-20.00Ω in steps of 10mΩ
0-200.0Ω in steps of 100mΩ
0-2000Ω in steps of 1Ω
Accuracy:
±0.5% of reading ±2 digits over the operating temperature range,
-15ºC to +55ºC, with the supplied test leads.
Test current:
1mA=>2000Ω range
10mA=>200 / 20Ω ranges
100mA=>2000m / 200mΩ ranges
Test current accuracy: ±0.1%.
Protection fuses:
Mains = 0.5A, HBC, 5 × 20mm, DIN
Current = 0.5A, HBC, 5 × 20mm, DIN
Voltage = 0.5A, HBC, 5 × 20mm, DIN
Maximum output voltage(C1~C2): 20V.
Dimensions: 250(L) × 190(W) × 110(D)mm.
Weight: 1500g.
Accessories:
Instruction manual,
Test leads,
Shoulder belt,
Batteries: 1.5V(AA) × 8
Safety Standard
EN 61010-1 CAT IV 20V
EN 61326-1